Marquest Scientific, Inc.
2950 Airway Ave., Bldg. C-5
Costa Mesa,CA,92626,US
Phone: 714-491-9191
Fax: 714-491-9199
Email: info@marquestscientific.com
Website: http://www.marquestscientific.com
All Categories > Laboratory Fixtures - PVC, Natural Polypro, & High Purity PVDF > DURALINEā„¢ DI Water Laboratory Faucets (Ultra Durable) > Recirculating Duraline DI Water Lab Faucets (Deck Mount) > View Items  

Recirculating Duraline DI Water Lab Faucets (Deck Mount)

Check up to five results to perform an action.

DURALINE™ Recirculating Laboratory Faucets — Natural Polypropylene (PPN) or PVDF
How it works (core): For return-to-tank DI/UPW loops using a non-pressurized reservoir. A factory-installed compact recirculating module sits below the deck at the inlet of the thru-deck nipple shank. Module ports are 3/8" FNPT inlet from the supply loop and 3/8" FNPT outlet back to the tank. A 0.125" ID recirculating tube seats inside the module, exits the outlet, and returns water to the reservoir to keep the loop moving when the dispense valve is closed.

Materials & purity (wetted path):
Natural PP or PVDF; Virgin PTFE seals; no elastomers or lubricants used in construction; non-metallic path; zero-dead-leg at dispense.

Ports & tube interface:
Module 3/8" FNPT inlet and 3/8" FNPT outlet; accepts 0.125" ID recirculating tube that returns to the non-pressurized tank.

Key specs:
Target velocity ≥ 4.5 ft/s (with 0.125" ID tube, about 0.2 GPM ≈ 5.2 ft/s); pressure rating 120 PSI.

Valve & control:
DURALINE control valve with positive shutoff and 120° metering; encapsulated PTFE shutoff seal kept in compression for constant on/off duty; PTFE stem seal; renewable cartridge enables in-place service.

Models:
DL-DDR-2 (Natural PP) · DL-DDR-3 (Natural PVDF).

Applications:

  • DI/UPW benches requiring continuous return-to-tank circulation between dispenses
  • High-purity labs mitigating stagnation and biofilm risk at the point of use
  • Multi-faucet series loops with a common return to a non-pressurized reservoir

Series piping note:
For faucets in series, use a common return and introduce a small downstream ΔP (diverter tee, valve, or orifice plate) at each faucet to ensure positive recirculation while maintaining dispense pressure.

Installation notes:
Use like materials across components; follow best-practice joining for PP or PVDF; verify velocity at the recirculating tube and set the required ΔP when piping in series.

Next step:
Select an Item # below for full specifications, installation details, and downloads.





  Results 1 - 2 of 2 1 

Item #

Materials

Configuration

Inlet x Outlet

Max PSI

List Price

DL-DDR-2 Natural Polypropylene Deck Mnt 3/8" Fem NPT 150 $720.50
DL-DDR-3 PVDF Deck Mnt 3/8" Fem NPT 150 $794.50
  Results 1 - 2 of 2 1 
Print Back